Verification report 01 · Parallel comparison

105 milk samples against a Bentley flow cytometer

A parallel comparison on fresh Holstein milk spanning 21,000 to 2,310,000 cells/mL. Both instruments measured every sample from the same tube, on the same day, back to back.

0.9955
Pearson r
0.9995
Regression slope
96.2%
Within 95% limits
97.1%
Agreement at 200k
Design

What was measured, and how

ItemDetail
Samples105 fresh Holstein milk samples, collected from commercial herds
Concentration range21,000 – 2,310,000 cells/mL
Reference instrumentBentley flow cytometer, in routine laboratory use
Sample handlingSame tube for both instruments, warmed to 40 °C for 10 minutes, mixed immediately before sampling
SequenceBoth instruments run back to back on each sample, not in separate sessions
Statistics reportedPearson r, Spearman ρ, regression slope and intercept with 95% confidence intervals, mean bias, 95% limits of agreement, bias by concentration band, classification agreement and Cohen's kappa at 200k and 400k
Result 1

The fit sits on the identity line

A slope of 0.9995 means no proportional bias across the whole range. This is the one comparison of the three where that holds.

Scatter plot of CytoVue SCC against a Bentley flow cytometer over 105 milk samples, with regression line and line of identity, r 0.9955
CytoVue SCC against the Bentley reference, n = 105. Solid line is the regression y = 11.95 + 0.9995x; dashed line is y = x. Both axes in thousands of cells/mL.
StatisticValue95% confidence interval
Pearson r0.9955—
Coefficient of determination R²0.9911—
Spearman ρ0.9954—
Regression slope0.99950.9810 – 1.0181
Regression intercept+11.95 k/mL1.95 – 21.95

The slope confidence interval contains 1.0, so there is no evidence of proportional bias. The intercept interval does not contain zero: the instrument reads about 12,000 cells/mL above the reference across the whole range, which is a fixed offset rather than a scaling error.

Result 2

Where the two instruments disagree

A fixed offset of 12,000 cells/mL is negligible at 2,000,000 and decisive at 50,000. The bias plot shows the whole distribution, not just its average.

Bland-Altman plot of CytoVue SCC minus Bentley reference against the mean of the two methods, mean bias plus 11.8, limits of agreement minus 58.7 to plus 82.2
Bland–Altman plot, n = 105. Solid line is the mean bias; dashed lines are the 95% limits of agreement. Vertical axis in thousands of cells/mL.
StatisticValue
Mean bias (CytoVue − Bentley)+11.76 k/mL
Standard deviation of the bias35.94 k/mL
95% limits of agreement−58.68 to +82.21 k/mL
Samples inside the limits101 of 105 — 96.2%

96.2% inside the limits is above the 95% that the agreement criterion asks for. The four samples outside are visible on the plot: three sit above the upper limit at mid-range concentrations, one below the lower limit at 1,600,000 cells/mL.

Bias by concentration band

Absolute bias is roughly constant, so relative bias falls as concentration rises. Below 100,000 cells/mL the count events are few enough that Poisson noise dominates, and a 20% relative difference at that level is expected of any counting method — including between two reference analysers.

Bar chart of relative bias by concentration band: plus 22.8 percent below 100 thousand, plus 4.8 percent from 100 to 500 thousand, plus 2.0 percent from 500 to 1000 thousand, plus 2.8 percent above 1000 thousand
Mean relative bias by reference-value band, n = 105.
Reference band (cells/mL)Mean relative bias
Below 100,000+22.8%
100,000 – 500,000+4.8%
500,000 – 1,000,000+2.0%
Above 1,000,000+2.8%
Result 3

Repeatability inside the comparison

Each sample was measured more than once, so the comparison also yields a repeatability distribution on real milk rather than on a spiked control.

Histogram of per-sample coefficients of variation across 105 milk samples, with 41.9 percent at or below 5 percent and 88.6 percent at or below 10 percent
Distribution of the coefficient of variation across the 105 samples.
CV bandShare of samplesCumulative
≤ 5%41.9%41.9%
5 – 10%46.7%88.6%
10 – 15%10.5%99.1%
15 – 20%1.0%100%

88.6% of samples repeat within 10%. The tail above 15% is concentrated at the low end of the range, where a handful of cells moves the percentage.

Four channels, one instrument

A four-channel chip is only useful if the four channels agree. The deviation of each channel from the sample mean is smaller than the difference between the two instruments, which is what should be expected: the channels share one optical path and one calibration.

ChannelMean deviation from the sample mean
Channel 1+3.14%
Channel 2−0.22%
Channel 3−1.19%
Channel 4−1.73%
Result 4

Do the two instruments classify the same sample the same way?

This is the number that matters commercially. A milk payment or a herd decision is made against a threshold, not against a correlation coefficient.

ThresholdBoth agree highBoth agree lowCytoVue high, reference lowCytoVue low, reference highAgreementCohen's kappa
200,000 cells/mL13890397.1%0.939
400,000 cells/mL10930298.1%0.958

No sample was classified high by CytoVue and low by the reference at either threshold, so the disagreement is one-directional: CytoVue classified three samples as low at 200,000 that the reference placed above it. Kappa above 0.9 is conventionally read as almost perfect agreement.

Ask for the full Bentley comparison

The complete report carries the per-sample table for all 105 samples, the regression diagnostics and the channel-level data. Tell us what reference instrument you run and we will send the matching document.