The evidence behind the numbers
Four documents. Two are parallel comparisons against imported reference analysers, run on real milk from real farms. One is a test a customer ran independently on their own DeLaval instrument. The fourth is the protocol we recommend you use to check any somatic cell counter — ours included — before you commit to one.
Pick the one that matches your situation
If you already run a reference analyser, the two parallel comparisons tell you what to expect. If you do not, the self-test protocol tells you how to evaluate without one.
Against a Bentley flow cytometer
105 fresh Holstein milk samples, 21,000 to 2,310,000 cells/mL, measured on both instruments from the same tube.
- r
- 0.9955
- Slope
- 0.9995
- Within limits
- 96.2%
Against a FOSS CombiFoss, at a DHI laboratory
Run at the customer's own DHI centre in Hohhot, against the FOSS analyser they use every day. Two CytoVue units, four chambers each.
- r
- 0.9981
- Threshold agreement
- 100%
- CV at 500k
- 2.65%
Against a DeLaval analyser, by the customer
We did not run this one. A customer tested it against their own DeLaval instrument and also timed the workflow and the stability of stained samples.
- r
- 0.9926
- 8 samples
- ~14 min
- Stain stable
- 25 min
How to test a somatic cell counter before you buy one
Four measurements that decide whether an instrument is worth its price, the acceptance thresholds we would hold ourselves to, and six ways an evaluation goes wrong.
- Items
- 4
- Criteria
- 12
- Pitfalls
- 6
Four studies, side by side
The same statistics reported the same way, so you can compare them without opening each document.
| Study | Samples | Reference instrument | r | Mean bias | Threshold agreement |
|---|---|---|---|---|---|
| Bentley parallel comparison | 105 | Bentley flow cytometer | 0.9955 | +11.8 k/mL | 97.1% at 200k 98.1% at 400k |
| FOSS on-site comparison | 17 | FOSS CombiFoss FT+300 | 0.9981 | +0.6 k/mL | 100% at 200k and 400k |
| Independent customer test | 12 | DeLaval analyser | 0.9926 | see report | not reported |
| Self-test protocol | — | — | — | — | — |
Correlation coefficients of this size are what a working instrument looks like. They are not the whole story: a slope below 1 means the instrument reads low at the top of the range, and the two comparisons show exactly that. Each report states where the difference sits rather than only quoting r.
How these numbers were produced
Samples come from real herds
Every comparison used fresh milk from commercial Holstein herds, not spiked buffer or reconstituted standards. Samples were collected, transported chilled and measured the same day. A milk matrix has fat globules outnumbering somatic cells by thousands to one, and any instrument that has not been run on real milk has not been tested.
Same tube, back to back
Both instruments measured from the same tube, warmed to 40 °C for ten minutes and mixed before sampling. The two instruments were run alternately rather than in two separate sessions, so a change in the sample between sessions cannot be mistaken for a difference between instruments.
Bias is reported, not hidden
A correlation coefficient alone can hide a proportional offset: an instrument reading exactly 1.10 times the reference still produces an r above 0.99. Each report therefore gives the regression slope and intercept, the mean bias, the 95% limits of agreement and the bias broken down by concentration band.
The decision threshold matters more than r
Milk is bought and rejected against a threshold, usually 200,000 or 400,000 cells/mL. What a buyer needs to know is not whether two numbers correlate but whether both instruments classify the same sample the same way. Each comparison reports classification agreement and Cohen's kappa for both thresholds.
Ask for the full reports
The published pages summarise each study. The complete documents, with the per-sample tables and the raw measurement series, are available on request — tell us which reference instrument you run and we will send the matching one.