CytoVue SCC — Milk Somatic Cell Counter
Fluorescence staining, microscopy imaging and fully automatic counting. A number you can put in front of a supplier, produced in 90 seconds for four samples, at a cost per test that makes daily testing affordable.

What the instrument actually measures
Raw milk is stained with a fluorescent nucleic acid dye, imaged through a 4× fluorescence objective and counted automatically. That is the same counting principle the large reference analysers in a dairy laboratory use — the difference is the size of the box it fits in.
Correlation with FOSS and Bentley reference instruments is better than 0.98, and both models share one optical path, one algorithm and one test chip. A figure produced in a parlour sits on the same chart as one produced in a laboratory.
The method suits cow milk and equally goat, buffalo and camel milk, because it counts stained nuclei rather than measuring a bulk property that shifts with fat and protein content.
How the count is produced → The optical path, the three counting steps, the standard the method belongs to, and the four conditions under which the number stops being reliable.

Why CytoVue SCC
Six points that decide a purchase in a milk laboratory, at a collection point, or on a farm.
Accuracy you can put in a contract
CV ≤ 7% at 200,000 cells/mL and CV ≤ 5% at 500,000, with r > 0.98 against FOSS and Bentley. The precision is stated at the concentrations where a payment decision is actually made.
The image is kept, not just the number
Every result is stored with the original stained image. When a supplier disputes a rejection there is something to look at, not just a figure on a printout.
Five steps, ninety seconds
Mix, warm, pipette, load, read. No dilution, no incubation, no plate reader, no separate wash cycle.
3 kg on a bench, 2.7 kg in a hand
The benchtop unit runs from a 12 V supply; the portable one carries its own battery and fits in a carry case, so testing can happen in a parlour instead of a sample bag going to a laboratory.
A cost per test that allows daily testing
Chips and reagent are made in-house, so the per-sample cost does not carry a third party's margin. Monitoring only works if it is cheap enough to run every day.
Support that answers within 24 hours
One year of warranty, a spare instrument in mainland China while yours is repaired, and maintenance for the life of the unit.
How it compares → The three methods used to count somatic cells, and what each one costs you in accuracy, evidence and time.
SCC100 and SCC200, side by side
Four hardware differences and one display difference. Optics, algorithm and consumable are identical.


| SCC100 | SCC200 | |
|---|---|---|
| Form factor | Benchtop | Portable |
| Display | 7-inch capacitive touchscreen | 4.3-inch touchscreen |
| Power | External DC 12 V / 3.0 A | Internal lithium battery |
| Weight | 3 kg | 2.7 kg |
| Dimensions | 230 × 230 × 245 mm | 135 × 243 × 200 mm |
| Image review | Original stained image with annotation | Original stained image, no annotation |
| Typical setting | Laboratory, milk station, third-party testing | Farm, collection point, on-site testing |
What both models do identically
These figures apply to SCC100 and SCC200 without distinction. They are the reason a result from a farm unit and a result from a laboratory unit can be plotted on the same chart.
| Parameter | Value |
|---|---|
| Measuring principle | Fluorescence staining + microscopy imaging + automatic counting |
| Measuring range | 0 – 5,000,000 cells/mL |
| Detection time | ≤ 90 seconds for 4 samples |
| Throughput | 80 tests/hour |
| Correlation | r > 0.98 against FOSS / Bentley reference instruments |
| Repeatability | CV < 7% at 200,000 cells/mL; CV < 5% at 500,000 cells/mL |
| Dilution required | None |
| Cell image | Yes — stained image stored with every result; annotation on SCC100 |
| Sample volume | 100 µL taken; 8.5 µL loaded per channel |
| Effective measured volume | 4.5 µL per channel |
| Stain | SytoG, dry-coated inside the test chip |


Verification and comparison reports → Parallel comparisons against Bentley and FOSS analysers, an independent customer test against DeLaval, and the protocol we recommend you use to check us.
Five steps from sample to number
No dilution, no incubation, no separate wash cycle.
Take and mix the milk
Bulk tank samples are stirred before drawing; individual cow samples discard the first two squirts.
Warm to 40 °C
The sample is brought to 40 °C in a water bath so the fat is liquid and the sample is homogeneous.
Stain in the reagent tube
100 µL of milk is pipetted up and down in the supplied stain tube; nuclei are fluorescently labelled.
Load 8.5 µL onto the chip
One channel of a four-channel CytoChip, which is used once and discarded.
Read the result
The instrument images and counts the channel, and reports cells/mL, the grade and the image, in 90 seconds for four samples.

The same five steps as a single reference strip — useful to print and pin above the bench.
The whole procedure on an SCC100, filmed end to end: warming to 40 °C, staining in the reagent tube, loading 8.5 µL per channel, and four results in 90 seconds. English narration.
Where the count gets used
Raw milk changes hands several times before it is processed. The somatic cell count decides the price at every one of those handovers.
Dairy farm — herd and individual cow
Weekly bulk tank readings show the trend; individual quarter samples identify the cow behind a rise. A portable unit lets the check happen in the parlour instead of a sample bag going to a laboratory.
Milk processor — raw material intake
Incoming milk is graded on arrival. A benchtop unit in the receiving laboratory produces the figure that goes on the intake record, with the image kept as evidence.
Third-party testing laboratory
Sampling rounds and market inspections need a result at the point of sampling, not a week later. Battery operation removes the dependency on a power supply at the site.
Research and teaching
Showing the stained image and explaining how nuclei are identified is part of the work; the larger annotated view on the benchtop model is the reason to choose it here.
What to order
One instrument, one chip type, one reagent. A site running both models does not have to stock two consumable lines.
| Item | Description |
|---|---|
| CytoVue SCC100 | Benchtop unit, 7-inch capacitive touchscreen, external DC 12 V / 3.0 A supply, image annotation |
| CytoVue SCC200 | Portable unit, 4.3-inch touchscreen, internal lithium battery, 2.7 kg |
| CytoChip | Single-use test chip, 4 channels, 96 chips per box (384 tests) |
| Stain reagent | SytoG dry-coated in the chip; reagent pouches supplied with each box of chips |
What happens after the instrument arrives
An instrument that stops working during harvest is a loss, not an inconvenience. These are the arrangements that keep it running.
| Item | What is included |
|---|---|
| Warranty | 1 year on the main unit, parts and labour, worldwide |
| Spare instrument | Provided in mainland China while your unit is being repaired. Outside mainland China, contact your local distributor — the loan unit and the repair are arranged and carried out by them. |
| Maintenance | Lifetime maintenance support |
| Response time | Technical support answers within 24 hours |
Frequently asked
0 to 5,000,000 cells/mL and 90 seconds for four samples on one chip, which is 80 tests per hour. A heavily infected sample does not need diluting before it goes on — the top of the range is twelve times the EU limit of 400,000 cells/mL.
Fluorescence staining, microscopy imaging, automatic counting. Nuclei are stained with a nucleic acid dye, a 4× fluorescence objective images the channel field by field, and software detects the stained nuclei and converts the count to cells/mL using the measured volume. The full principle.
No. 100 µL of raw milk goes into the stain tube, is pipetted up and down to mix, and 8.5 µL of the stained sample is loaded onto the chip. No dilution series, no incubation, no reagent preparation.
Four channels per chip, so four samples per chip. 100 µL of milk is taken per sample, 8.5 µL is loaded per channel, and the effective measured volume is 4.5 µL per channel.
No. CytoChip is single use. One box holds 96 chips, which is 384 tests, and a box is normally paired with four stain reagent pouches.
Instruments and consumables are both made in-house, so the cost per test does not carry a third party's margin. Send us your tests per day and we will quote the instrument and the annual consumable quantity together.
If both results sit inside the tolerance the method allows, neither is wrong. NY/T 800-2004 permits a relative difference of up to 15% between two parallel determinations in the same laboratory, and cross-method comparisons are read at that order of magnitude. When the gap is clearly wider, check sampling and sample storage before suspecting the instrument.
One year on the main unit. In mainland China a spare instrument is provided while yours is being repaired; overseas, one year of parts and labour. Maintenance continues for the life of the instrument and support answers within 24 hours.
Request a quote for CytoVue SCC
Tell us the sample volume per day and where the instrument will sit. We will confirm the model and the consumable quantity.
