Verification report 02 · On-site comparison

17 samples at a DHI laboratory, against a FOSS CombiFoss

Run on site at a customer's own DHI centre in Hohhot, against the FOSS analyser the laboratory uses every day for routine payment testing. Two CytoVue units took part, four chambers per sample on each.

0.9981
Pearson r, unit 1
100%
Agreement at 200k and 400k
2.65%
CV at 500k, 4-chamber mean
1.000
Cohen's kappa
Design

What was measured, and how

This comparison was requested by the customer, not proposed by us. The reference instrument, the samples and the laboratory were theirs.

ItemDetail
SiteCustomer DHI centre, Hohhot, Inner Mongolia
Samples17 Holstein milk samples, 21,000 – 3,169,000 cells/mL
Reference instrumentFOSS CombiFoss FT+300, in daily use for payment testing
Instruments under testTwo CytoVue units, four chambers per sample on each
Sample handlingSame tube, warmed to 40 °C for 10 minutes, mixed immediately before sampling
DateMay 2025
Result 1

Correlation is high, and the slope is not 1

Both facts are in the report. Quoting only the first would be misleading.

Scatter plot of CytoVue SCC against a FOSS CombiFoss over 17 milk samples, with regression line and line of identity, r 0.9981
CytoVue SCC against the FOSS CombiFoss reference, n = 17. Solid line is the regression y = 39.1 + 0.954x; dashed line is y = x. Both axes in thousands of cells/mL.
StatisticUnit 1Unit 2
Pearson r0.99810.9985
Coefficient of determination R²0.99610.9970
Spearman ρ0.99750.9975
Regression slope0.95380.9248
Slope 95% confidence interval0.9210 – 0.98660.8969 – 0.9528
Regression intercept+39.1 k/mL+45.2 k/mL
Mean bias against FOSS+0.64 k/mL−17.37 k/mL
Samples inside the 95% limits94.1%88.2%

Both slope intervals lie entirely below 1.0. Against this FOSS instrument the CytoVue reads progressively lower as concentration rises: at 3,000,000 cells/mL the gap is roughly 5% on unit 1 and 8% on unit 2. The mean bias is near zero because the positive intercept and the slope below 1 cancel each other out across the range — which is exactly why a single average bias figure is not enough on its own.

Result 2

Repeatability, measured on real milk

Two concentration levels, each measured on seven chips, four chambers per chip. The distinction between single-chamber and four-chamber-mean CV is the one that matters in practice.

Around 500,000 cells/mL

Counting basisResultsMeanStandard deviationCV
Each chamber counted separately28477.0 k/mL18.4 k/mL3.86%
Four-chamber mean as the result7477.0 k/mL12.6 k/mL2.65%

Around 200,000 cells/mL

Counting basisResultsMeanStandard deviationCV
Each chamber counted separately24191.7 k/mL13.2 k/mL6.88%
Four-chamber mean as the result6191.7 k/mL7.5 k/mL3.90%

Reporting the four-chamber mean rather than a single chamber cuts the CV by roughly a third at both levels. This is the same averaging effect that makes any replicate measurement better than a single one, and it is why the instrument reports the mean by default.

Result 3

Classification agreement was perfect

17 samples is a small set, but at both payment thresholds the two methods put every sample in the same class.

ThresholdBoth highBoth lowDisagreementAgreementCohen's kappa
200,000 cells/mL1340100%1.000
400,000 cells/mL1070100%1.000

Against the Bentley reference the same test gave 97.1% and 98.1%. Perfect agreement on 17 samples is a weaker claim than 97% on 105 — with this sample size a single disagreement would have moved the figure by six points — but it does establish that the slope difference does not translate into a classification difference at the thresholds that decide payment.

Result 4

Two units, same sample

Two instruments of the same model, on the same samples, in the same session. The difference between them sets the floor for any comparison against a third instrument.

StatisticUnit 1 − Unit 2
Mean bias+18.01 k/mL
Relative bias+2.16%
Standard deviation of the bias31.98 k/mL
Samples inside the 95% limits94.1%

Two units of the same model differ by about the same amount as either unit differs from the reference. That is the honest reading of a comparison like this: part of what looks like instrument-versus-reference difference is instrument-versus-instrument variation, and no single unit can be treated as the definition of correct.