17 samples at a DHI laboratory, against a FOSS CombiFoss
Run on site at a customer's own DHI centre in Hohhot, against the FOSS analyser the laboratory uses every day for routine payment testing. Two CytoVue units took part, four chambers per sample on each.
What was measured, and how
This comparison was requested by the customer, not proposed by us. The reference instrument, the samples and the laboratory were theirs.
| Item | Detail |
|---|---|
| Site | Customer DHI centre, Hohhot, Inner Mongolia |
| Samples | 17 Holstein milk samples, 21,000 – 3,169,000 cells/mL |
| Reference instrument | FOSS CombiFoss FT+300, in daily use for payment testing |
| Instruments under test | Two CytoVue units, four chambers per sample on each |
| Sample handling | Same tube, warmed to 40 °C for 10 minutes, mixed immediately before sampling |
| Date | May 2025 |
Correlation is high, and the slope is not 1
Both facts are in the report. Quoting only the first would be misleading.

| Statistic | Unit 1 | Unit 2 |
|---|---|---|
| Pearson r | 0.9981 | 0.9985 |
| Coefficient of determination R² | 0.9961 | 0.9970 |
| Spearman ρ | 0.9975 | 0.9975 |
| Regression slope | 0.9538 | 0.9248 |
| Slope 95% confidence interval | 0.9210 – 0.9866 | 0.8969 – 0.9528 |
| Regression intercept | +39.1 k/mL | +45.2 k/mL |
| Mean bias against FOSS | +0.64 k/mL | −17.37 k/mL |
| Samples inside the 95% limits | 94.1% | 88.2% |
Both slope intervals lie entirely below 1.0. Against this FOSS instrument the CytoVue reads progressively lower as concentration rises: at 3,000,000 cells/mL the gap is roughly 5% on unit 1 and 8% on unit 2. The mean bias is near zero because the positive intercept and the slope below 1 cancel each other out across the range — which is exactly why a single average bias figure is not enough on its own.
Repeatability, measured on real milk
Two concentration levels, each measured on seven chips, four chambers per chip. The distinction between single-chamber and four-chamber-mean CV is the one that matters in practice.
Around 500,000 cells/mL
| Counting basis | Results | Mean | Standard deviation | CV |
|---|---|---|---|---|
| Each chamber counted separately | 28 | 477.0 k/mL | 18.4 k/mL | 3.86% |
| Four-chamber mean as the result | 7 | 477.0 k/mL | 12.6 k/mL | 2.65% |
Around 200,000 cells/mL
| Counting basis | Results | Mean | Standard deviation | CV |
|---|---|---|---|---|
| Each chamber counted separately | 24 | 191.7 k/mL | 13.2 k/mL | 6.88% |
| Four-chamber mean as the result | 6 | 191.7 k/mL | 7.5 k/mL | 3.90% |
Reporting the four-chamber mean rather than a single chamber cuts the CV by roughly a third at both levels. This is the same averaging effect that makes any replicate measurement better than a single one, and it is why the instrument reports the mean by default.
Classification agreement was perfect
17 samples is a small set, but at both payment thresholds the two methods put every sample in the same class.
| Threshold | Both high | Both low | Disagreement | Agreement | Cohen's kappa |
|---|---|---|---|---|---|
| 200,000 cells/mL | 13 | 4 | 0 | 100% | 1.000 |
| 400,000 cells/mL | 10 | 7 | 0 | 100% | 1.000 |
Against the Bentley reference the same test gave 97.1% and 98.1%. Perfect agreement on 17 samples is a weaker claim than 97% on 105 — with this sample size a single disagreement would have moved the figure by six points — but it does establish that the slope difference does not translate into a classification difference at the thresholds that decide payment.
Two units, same sample
Two instruments of the same model, on the same samples, in the same session. The difference between them sets the floor for any comparison against a third instrument.
| Statistic | Unit 1 − Unit 2 |
|---|---|
| Mean bias | +18.01 k/mL |
| Relative bias | +2.16% |
| Standard deviation of the bias | 31.98 k/mL |
| Samples inside the 95% limits | 94.1% |
Two units of the same model differ by about the same amount as either unit differs from the reference. That is the honest reading of a comparison like this: part of what looks like instrument-versus-reference difference is instrument-versus-instrument variation, and no single unit can be treated as the definition of correct.
Ask for the full FOSS comparison
The complete report carries the per-sample four-chamber data for both units and the full regression diagnostics. If you run a CombiFoss and want the version that matches your instrument, tell us and we will send it.