Twelve samples, tested by the customer on their own DeLaval analyser
We did not run this test and we did not choose the samples. A customer evaluated the instrument against their own DeLaval analyser and, while they were at it, timed the workflow and checked how long a stained sample stays valid. Their findings are published here as they reported them.
Agreement with the customer's DeLaval instrument
Twelve samples, measured on both instruments. The customer's own data, plotted without adjustment.

| Statistic | Value |
|---|---|
| Samples | 12 |
| Pearson r | 0.9926 |
| Coefficient of determination R² | 0.9853 |
| Regression slope | 0.898 |
| Regression intercept | +15.4 k/mL |
The direction of the slope matches the FOSS comparison: below 1.0, meaning the CytoVue reads lower than the reference as concentration rises. Twelve samples is enough to see the trend and not enough to quantify it — the confidence interval on a slope from twelve points is wide. Read this report as a signal that agrees with the larger study, not as a measurement in its own right.
What the customer also measured
These four tests were not part of the comparison. They came out of a customer trying to work out how the instrument would behave in their own routine, and they answer questions a correlation coefficient never touches.
A stained sample stays valid for about 25 minutes
Two samples were stained, held at room temperature and re-read. Over 25 minutes the readings moved by less than 1%, which is inside the repeatability of the instrument itself.
| Time after staining | Sample 1 | Sample 2 |
|---|---|---|
| Immediate | 30.09 k/mL | 53.77 k/mL |
| 25 minutes | 31.03 k/mL | 54.09 k/mL |
A loaded chip does not keep
The same two samples were loaded onto chips and left standing before reading. Both drifted down steadily, by 15% and 25% over 55 minutes. The mechanism is straightforward: cells settle out of the suspension in the channel, so fewer pass through the imaging field. Load the chip and read it.
| Standing time after loading | Sample 1 | Sample 2 |
|---|---|---|
| 0 minutes | 28.62 k/mL | 21.18 k/mL |
| 5 minutes | 27.67 k/mL | 19.05 k/mL |
| 30 minutes | 26.21 k/mL | 16.67 k/mL |
| 55 minutes | 24.27 k/mL | 15.99 k/mL |
Eight samples in one run takes about fourteen minutes
The customer compared running eight samples in a single session against running two, and found the readings unchanged. Batching does not cost accuracy, which is what makes the throughput figure usable rather than theoretical.
| Mode | Sample 1 | Sample 8 |
|---|---|---|
| Eight samples in one run | 47.28 k/mL | 21.18 k/mL |
| Two samples in a single run | 46.38 k/mL | 19.03 k/mL |
A cold sample reads high
The same sample measured straight from the refrigerator and again after ten minutes in a 40 °C water bath. The cold sample read higher on both, by 14% and 10%. Fat is still crystallised when cold and scatters differently, and the cells have not been released from the cream layer. This is why warming is written as a mandatory step rather than a recommendation.
| Condition | Sample 1 | Sample 2 |
|---|---|---|
| Cold, measured directly | 33.86 k/mL | 97.90 k/mL |
| 40 °C water bath, 10 minutes | 38.57 k/mL | 88.89 k/mL |
Sample 1 read lower cold and sample 2 read higher cold. With two samples this is not a rule, it is a demonstration that an unwarmed sample is unpredictable. Treat the step as compulsory and the question does not arise.
A small test that is worth publishing anyway
What it establishes
That an independent user, using their own reference instrument and their own samples, reproduced the correlation the larger studies report, and that the operational behaviour of the instrument — how long a stained sample lasts, what happens to a loaded chip, what warming does — is predictable enough to write into a standard operating procedure.
What it does not establish
Twelve samples cannot pin down a slope. The 0.898 figure should be read alongside the 0.9995 from the Bentley study and the 0.954 from the FOSS study, not instead of them. Three studies pointing the same way is evidence; one study of twelve samples is a data point.
Test it against your own reference instrument
If you run a DeLaval, FOSS or Bentley analyser in your own laboratory, we will send a unit and the sample plan. You keep the data and publish the result, whichever way it falls.